Special 510(k) Stretch: Indications-for-Use Changes in 428 Software Clearances An analysis of 486 Special 510(k) clearances for software as a medical device from 1998 to 2026 found that 24 of the 428 clearances with comparable predicate indications text (5.6%) scored 5 or higher on a 0-10 AI-model panel, indicating a material expansion of indications for use, while 250 (58%) made no change and 118 (28%) made a trivial one. The share scoring 5 or higher rose from 3.4% before FDA's September 13, 2019 Special 510(k) guidance to 8.4% after it (Fisher's exact test, p = 0.03), and no clearance scored above 7. In three of the four largest changes, including Annalise Enterprise (K250831) and Stethophone Pro (K252595), the manufacturer already held the added indications under another of its own clearances. A Special 510 k is a manufacturer's change to its own cleared device, with a 30-day review target. FDA's 2019 guidance https://fda.innolitics.com/guidance/special-510k-program?v=gfc09bf3b909-72dfe4d1f679a1bb states that certain changes to the indications for use may go through this pathway. This article measures how far the indications for use moved in every Special 510 k for software as a medical device SaMD in the FDA Device Explorer https://fda.innolitics.com/ . The cohort holds 486 clearances from 1998 to 2026. For 428 of them, the cited predicates have comparable indications text. A panel of AI models scored each change from 0 to 10 against the indications of the cited predicates. Of the 428 clearances, 250 58% make no change and 118 28% make a trivial one. Twenty-four 5.6% score 5 or higher, which means a material expansion on at least one dimension. No clearance scores above 7. In three of the four largest changes, the manufacturer already held the added indications under another of its own clearances. Annalise Enterprise