How Far Can a 510(k) Predicate Stretch? One in five AI/ML 510(k) clearances cites a predicate under a different product code, with 85 devices predicated on a different body region, 35 swapping imaging modality, and 122 crossing 21 CFR regulation numbers, according to an analysis of 1,457 AI/ML-flagged records in the Innolitics FDA Device Explorer. The analysis found brain triage software predicated abdominal triage, a desktop urinalysis analyzer predicated a phone camera test strip app, and Aidoc's head CT triage clearance became the predicate for cervical spine, abdominal free gas, and rib fracture triage, with no outcome data required. Ask most people what a valid predicate looks like and you get the same checklist: same body region, same modality, same product code. The clearance record says otherwise. One in five AI 510 k s cites a predicate under a different product code. 85 cleared with a predicate targeting a different body region, 35 swapped imaging modality outright, and 122 predicate citations cross 21 CFR regulation numbers entirely. Brain triage software predicated abdominal triage. A desktop urinalysis analyzer was the predicate for an app that reads test strips with a phone camera. We measured how far a 510 k predicate can stretch, across every axis we could quantify, in 1,457 AI/ML-flagged records in the Innolitics FDA Device Explorer https://fda.innolitics.com/ . Six axes of predicate distance: anatomy, imaging modality, care setting and user, function class, product code, and regulation number. The first four come from criterion-based screening of every predicate-change reason in the corpus. The last two are computed mechanically: we resolved the product code of all 1,181 cited predicate records and mapped each to its 21 CFR regulation. Every K-number, DEN number, product code, and regulation in this article links to the primary record. We measured this two ways. Systematically pairing every AI clearance against every cited predicate and comparing the anatomical targets in their intended-use statements finds 85 devices about 1 in 15 whose predicate targeted a different specific body region, and 173 about 1 in 7 counting multi-region expansions and narrowings. The 85 devices that crossed outright are charted below, one primary crossing per device, and brain seeds more of them than any other region: 18 departures. The Aidoc arc shows what that means commercially. One head CT triage clearance became the predicate for cervical spine fracture K190896 https://fda.innolitics.com/device/K190896 , abdominal free gas K193298 https://fda.innolitics.com/device/K193298 , and rib fracture triage K202992 https://fda.innolitics.com/device/K202992 , four markets on one algorithm family. The stranger pairs cleared too. Cycle Clarity predicated ovarian follicle measurement on cardiac chamber segmentation K212012 https://fda.innolitics.com/device/K212012 , and Smart Soft predicated lumbar spine measurement on a brain volumetry tool K220497 https://fda.innolitics.com/device/K220497 . Nobody was asked for outcome data. The ask, over and over, was standalone performance in the new anatomy. One scoping note: anatomy here means body region, so the 85 counts only the major moves. Predicates that hop between organs inside the same region are far more numerous, and we did not capture them in this analysis. They are the subject of a separate article. This one tracks the documented big jumps. 95 modality-related crossings: 35 outright swaps, 27 additions where a second modality was bolted onto a cleared device, 16 where the sponsor crossed modality while citing a De Novo as the predicate BriefCase rode the ContaCT De Novo from CTA to non-contrast CT, K180647 https://fda.innolitics.com/device/K180647 , and 17 signal-source shifts among monitors and wearables. The swaps are the ones that surprise people. Mammography CADx predicated lung CT K251474 https://fda.innolitics.com/device/K251474 in 2025, and a CT-based planning tool predicated TEE ultrasound video K260300 https://fda.innolitics.com/device/K260300 the following year, so this is not some loophole FDA quietly closed. Contrast versus non-contrast CT behaves like a free move, with nine crossings running in both directions. And the cheapest expansion in the whole dataset is the modality addition, led by CT plus MR at 14 instances, because FDA treats the added modality as a validation domain rather than a new device. 34 care-setting crossings, and the direction is nearly one-way: specialist settings seeded 19, home and consumer absorbed 17. Clinician-overread ECG became independent consumer use K191406 https://fda.innolitics.com/device/K191406 . A desktop lab urinalysis analyzer became a patient's phone camera K210069 https://fda.innolitics.com/device/K210069 , which still strikes us as the most audacious predicate argument in the cohort, and it worked. Traffic occasionally runs backward too; TRIOS Dx brought a remote monitoring predicate into the dental clinic K260082 https://fda.innolitics.com/device/K260082 . One caveat: the first consumer device in each category went De Novo Apple ECG, DEN180044 https://fda.innolitics.com/device/DEN180044 , and everyone after rode it. 60 clearances crossed function classes, and quantification is the hub with 17 departures. One-rung hops are routine: quantification to CADx, segmentation to planning. Deliberate down-classing works and is underused: IQ-UIP repositioned a diagnostic classification predicate as passive notification K242467 https://fda.innolitics.com/device/K242467 . But multi-rung jumps into prognosis or autonomy are where De Novo risk concentrates, and that is the gate: every category-creating De Novo in the cohort ContaCT, QuantX, Caption Guidance, IDx-DR, ArteraAI Prostate introduced a claim type with no precedent anywhere. None was forced by anatomy, modality, code, or setting. 349 predicate citations cross product codes. The dominant flow is FDA recoding its own ecosystem: 150 citations migrate the legacy LLZ https://fda.innolitics.com/productcode/LLZ image-processing code into QIH https://fda.innolitics.com/productcode/QIH . Strip that wave out and 199 genuine crossings remain. Deeper still, 122 citations cross regulation numbers, densest inside cardiology's 870.x cluster, where ECG analysis, hemodynamic prediction, and cardiovascular quantification software have been borrowing each other's predicates for years. Pick your code by where the special controls and predicate pool are most favorable. The crossing will not be challenged; the comparison table will. | Axis | Cleared crossings | Headline finding | |---|---|---| | Anatomy | 85 outright 173 with multi-region shifts | Measured across every predicate pair; brain is the dominant seed and anatomy is the cheapest-looking big move | | Modality | 95 | Swaps cleared in every era; CT plus MR addition is the cheapest expansion; contrast versus non-contrast is near-free | | Care setting / user | 34 | Traffic runs outward toward home and consumer; category firsts were De Novos | | Function class | 60 | Quantification is the hub; one-rung hops are routine; multi-rung jumps ride De Novos | | Product code | 349 citations | One in five clearances cites a cross-code predicate | | Regulation number | 122 citations | Even 21 CFR boundaries are porous, densest in cardiology | Score a proposed device against its best candidate predicate in two steps. First the gate: does the device make a claim type with no cleared precedent in any anatomy or modality? If yes, plan for De Novo and decide whether to author the special controls deliberately. If the gate passes, score distance out of 100: | Axis | Weight | What the data shows | |---|---|---| | Output type and autonomy | 25 | Autonomy escalations forced De Novos; notification-only framing repeatedly bought cheaper clearances | | Decision consequence | 20 | Prognosis and risk-prediction claims all went De Novo first | | Modality | 15 | 55 cleared crossings; 5 points for a within-family variant, 15 for a full swap | | Use environment and user | 15 | Consumer-context firsts went De Novo; later entrants rode them | | Anatomy | 10 | 85 measured crossings including brain to abdomen | | Population shift | 10 | Population-only expansions cleared routinely | | Technology character | 5 | Rule-based to CNN swaps cleared silently; locked to adaptive cleared with a PCCP | Under 40, lead with a lean 510 k . Between 40 and 65, file with the highest-scoring axis neutralized in the substantial equivalence table. Over 65, accept De Novo or decompose the journey into two clearances: Clarius chained bladder K232257 https://fda.innolitics.com/device/K232257 to prostate K243853 https://fda.innolitics.com/device/K243853 to median nerve K250226 https://fda.innolitics.com/device/K250226 , and each hop read as a modest delta. | Axis | Transition | n | Example records | |---|---|---|---| | Anatomy | Heart → Vascular | 7 | | Artera ran the first-mover play with a straight face. They took prostate prognosis through De Novo in July 2025 DEN240068 https://fda.innolitics.com/device/DEN240068 , which cost them the long road: outcome-linked validation, novel special controls, the full category-creation tax. Ten months later, ArteraAI Breast cleared as a 510 k predicated on their own De Novo K254115 https://fda.innolitics.com/device/K254115 . They paid the toll once and then owned the bridge. Every follower in digital pathology prognosis now has to meet special controls Artera wrote. Clarius had a different problem: the distance from their starting point to their ambition was too far for one filing. So they walked. Bladder volume K232257 https://fda.innolitics.com/device/K232257 predicated prostate volume K243853 https://fda.innolitics.com/device/K243853 , while a tendon tool predicated median nerve segmentation K250226 https://fda.innolitics.com/device/K250226 . Three clearances in roughly three years, each a modest delta a reviewer could nod through, and each a sellable product while the next one was in flight. When a client's end-to-end plan scores as a stretch, this is usually our answer. Aidoc built a franchise on a single predicate root. One head CT triage clearance became the citation for cervical spine, abdominal free gas, and rib fracture, and later a quantification line, while competitors cited the same family to enter adjacent anatomies. The lesson cuts both ways. Your clearance is an asset you can compound, and it is also a doorway you may be opening for whoever files next. Yes, and not as an exotic exception: 85 cleared devices crossed body regions entirely, and moves between organs inside the same region are more numerous still. Hold modality, claim type, and output constant and the anatomy move is the cheapest big jump available. What FDA asked for, in clearance after clearance, was standalone performance in the new anatomy. Yes: 35 outright swaps cleared, from wrist radiographs predicating screening mammography K181704 https://fda.innolitics.com/device/K181704 to mammography CADx predicating lung CT. Full-distance swaps cleared most easily when the output type and claim stayed boring. Keep every other row of the comparison table identical. Routinely. One in five clearances cites a cross-code predicate, and 122 citations cross regulation numbers entirely. Pick your code by where the evidence burden is lightest, not by where the name fits. A granted De Novo is a fully valid predicate; 73 clearances cite one. ArteraAI Breast predicated the ArteraAI Prostate De Novo within ten months of its grant K254115 https://fda.innolitics.com/device/K254115 on DEN240068 https://fda.innolitics.com/device/DEN240068 . When your claim type has no cleared precedent anywhere: a new function class, a new autonomy level, or a new consumer context. If a device with your claim type exists in any organ or modality, you have a 510 k argument. No. Rule-based to CNN swaps cleared without ceremony K212516 https://fda.innolitics.com/device/K212516 , K212261 https://fda.innolitics.com/device/K212261 , and locked to adaptive cleared with a PCCP K243558 https://fda.innolitics.com/device/K243558 . Algorithm internals are the cheapest axis in the framework. 34 precedents say yes, and the traffic overwhelmingly runs toward the patient. If a consumer precedent exists in your category, ride it. If not, you are the De Novo. One-rung hops are routine: 60 cleared crossings. Deliberate down-classing also works: diagnosis repositioned as notification bought a faster review. Multi-rung jumps into prognosis or autonomy are where De Novo risk lives. Standalone performance in the new domain against a defensible reference standard, plus the usual software documentation. Distance on these axes changes what you validate, rarely how much clinic you must run. When the end-to-end distance scores too high, clear an intermediate device and predicate yourself. Clarius went bladder to prostate to median nerve in three years. Bunkerhill went coronary calcium K230223 https://fda.innolitics.com/device/K230223 to aortic valve K243229 https://fda.innolitics.com/device/K243229 to mitral annulus K250662 https://fda.innolitics.com/device/K250662 . The intermediate clearance is often marketable on its own. It is a position to exploit. The De Novo author writes the special controls every follower must meet, then converts each new organ into a cheap 510 k against its own device. Our data shows 25 same-target follow-ons drafting behind pioneers, so author special controls with followers in mind. Curated figures reflect the corpus as of August 8, 2026 1,457 AI records ; the systematic anatomy measurement reflects August 15 1,483 . The full study, six companion reports, and every dataset behind these tables are available from Innolitics; every citation above links to the free FDA Device Explorer https://fda.innolitics.com/ , where the underlying 510 k summaries are searchable in full text. Source: Innolitics FDA Device Explorer https://fda.innolitics.com/ , AI/ML-flagged cohort, N = 1,457 records as of August 8, 2026 1,483 by August 15, when the anatomy measurement ran , with Innolitics-extracted predicate-change reasons. Anatomy, modality, setting, and function datasets were curated by criterion-based screening; product-code and regulation crossings were computed against the full 179,331-record corpus and the openFDA classification database. The anatomy-crossing rate was additionally measured systematically: the primary anatomical target was extracted from the intended-use statement of every AI clearance and every cited predicate 2,061 records classified into eleven regions , and each subject-predicate pair was compared, yielding 85 strict region crossings and 173 including multi-region shifts among 1,267 clearances with a comparable anatomical predicate. We may have missed devices whose predicate reasons understate the change. The companion reports carry full tables, clickable transition matrices, and per-record links for each axis.