From Single-Finding Filings to Multi-Indication CADt: How FDA Triage AI Evolved (2018-2026) An Innolitics audit of all 132 FDA authorizations under 21 CFR 892.2080 with primary product codes QAS or QFM, covering 2018-02-13 through 2026-11-28, found that modular multi-indication bundling is becoming a repeatable regulatory pattern for radiological triage AI. The audit cites a2z Radiology AI's a2z-Omni-Triage (K262765), cleared in 100 calendar days on November 12, 2026 with a Predetermined Change Control Plan authorized, which brought the company to 26 cleared finding claims across three devices within 353 calendar days. The cohort comprised 1 De Novo and 131 510(k) authorizations, with Omni's nine-finding list and 26-claim total drawn from an Imaging Technology News report because no summary PDF was posted as of November 28, 2026. Download the 1-Page Regulatory Landscape Brief: Get the printable Evolution of Multi-Indication CADt Clearances 2018-2026 PDF https://innolitics.com/pdf/cadt-evolution-landscape-1pager.pdf with 34 clickable links to primary FDA summaries, guidance, and source reports. In November 2026, a2z Radiology AI cleared a2z-Omni-Triage K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 in 100 calendar days, bringing its abdomen-pelvis CT triage portfolio to 26 cleared finding claims across three devices https://highlight.innolitics.com/v/jLF0v46yMMDH h=1 in 353 calendar days between its first and third clearances. Less than a year earlier, when we published our teardown of a2z-Unified-Triage K252366 https://innolitics.com/articles/K252366/ , bundling seven abdominopelvic findings into a single dual-code 510 k looked like an outlier. Today, modular multi-indication bundling is emerging as a repeatable regulatory pattern across radiological triage AI. To measure how the category has shifted, we audited all 132 FDA authorizations with primary product code QAS or QFM under 21 CFR 892.2080 through November 28, 2026 using Innolitics FDA Device Explorer https://fda.innolitics.com N = 132 authorizations: 1 De Novo and 131 510 k s , extracting indication counts across every published summary in the nine-year cohort. Four empirical shifts stand out: K203508 K250831 K252366 K252970 K260906 K262765 K253578 Methodology and primary-source scope: Cohort records, dates, subsequent product codes, and whole-corpus indication extractions for all 132 primary-code 21 CFR 892.2080 authorizations 2018-02-13 through 2026-11-28 were pulled via Innolitics FDA Device Explorer https://fda.innolitics.com . Distinct-pathology counts distinguish multi-condition triage panels from single-pathology neurovascular segment lists K220709 , K251151 , K260729 , K261085 https://fda.innolitics.com/device/K261085?doc=summary&hl=10:Rapid%20VO%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20the%20CT%20Angiography%20%28CTA%29 , K261317 , aggregate ICH subtype lists K221456 , K241719 , K250694 , K262301 https://fda.innolitics.com/device/K262301?doc=summary&hl=9:Rapid%20ICH%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20non-enhanced%20head%20CT , and K203508 ’s non-clinical user-label update across 6 previously cleared BriefCase indications. FDA also assigned QAS or QFM as subsequent product codes on clearances filed under other primary codes, including K261368 https://fda.innolitics.com/device/K261368?doc=summary&hl=1:Trade%2FDevice%20Name%3A%20Rapid%20Platform%20%28R6.7%29&hl=1:Product%20Code%3A%20QIH%2C%20LLZ iSchemaView Rapid Platform R6.7, primary QIH with subsequent QAS and QFM and K262062 https://fda.innolitics.com/device/K262062?doc=summary&hl=1:Product%20Code%3A%20QBS%2C%20QFM%2C%20QIH&hl=2:also%20included%20the%20review%20and%20clearance%20of%20your%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29%3A%20Predetermined%20Change%20Control%20Plan%20CINA-Aneurysm Avicenna.ai CINA-Aneurysm, primary QBS with subsequent QFM and QIH , PCCP authorized , which sit outside this primary-code denominator. For K262765 https://fda.innolitics.com/device/K262765 , FDA’s 510 k record https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 confirms receipt on August 4, 2026, clearance on November 12, 2026 100 calendar days , primary QAS, subsequent QFM https://highlight.innolitics.com/v/pp1p85c14Mlw h=2 , and PCCP Authorized: Yes , with no summary PDF posted as of November 28, 2026; accordingly, Omni’s nine-finding list and 26-claim cumulative total come from the Imaging Technology News ITN report https://highlight.innolitics.com/v/jLF0v46yMMDH h=2 , and its PCCP scope remains unverified. Likewise, Aidoc’s Special 510 k K253265 https://highlight.innolitics.com/v/3r61d7uINGlI h=1 cleared November 6, 2025 is marked PCCP Authorized: Yes in FDA’s 510 k database without PCCP scope details in its posted summary. All other device counts, endpoints, cohorts, and PCCP scopes come directly from published FDA summaries indexed in Innolitics FDA Device Explorer. 21 CFR 892.2080 Computer-aided triage and notification CADt software is regulated under 21 CFR 892.2080 Class II with special controls . Under that regulation, FDA classifies authorizations across two product codes: QAS QFM Across the 132 primary-code authorizations by decision year, QAS accounts for 91 68.9% and QFM accounts for 41 31.1% . Annual authorization volume peaked at 24 authorizations in 2022 and settled to 15 to 19 per year since. What changed in 2025 and 2026 is indication density per clearance, dual-code QAS + QFM pairing, and PCCP adoption: | Decision Year | Total 21 CFR 892.2080 | Primary QAS | Primary QFM | Dual QAS + QFM | Confirmed PCCP | Median Review Days | |---|---|---|---|---|---|---| | 2018 | 3 | 3 | 0 | 0 | 0 | 137.0 | | 2019 | 9 | 5 | 4 | 0 | 0 | 102.0 | | 2020 | 16 | 12 | 4 | 0 | 0 | 132.0 | | 2021 | 10 | 5 | 5 | 0 | 0 | 125.5 | | 2022 | 24 | 15 | 9 | 0 | 0 | 116.5 | | 2023 | 19 | 11 | 8 | 1 | 0 | 133.0 | | 2024 | 18 | 12 | 6 | 0 | 0 | 146.5 | | 2025 | 18 | 15 | 3 | 2 | 2 | 100.5 | | 2026 through Nov 28 | 15 | 13 | 2 | 4 | 4 | 104.0 | | Total / Overall | 132 | 91 | 41 | 7 | 6 | 113.0 | Source: Innolitics FDA Device Explorer https://fda.innolitics.com 21 CFR 892.2080 primary QAS/QFM cohort, queried November 28, 2026 , supplemented with FDA’s live 510 k PCCP authorization flag for K253265 https://highlight.innolitics.com/v/3r61d7uINGlI h=1 and K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 . Review days reflect exact calendar days from receipt to decision including hold time . Confirmed PCCPs include 4 documented in published summaries K252366 https://fda.innolitics.com/device/K252366?doc=summary&hl=7:While%20the%20anatomical%20regions%20of%20interest%20differ%20%28abdomen%2Fpelvis%20for%20the%20subject%20device%20vs.%20head%20for%20the%20predicate%29%2C%20the%20core%20technology%2C%20intended%20use%20for%20triage%2C%20and%20principles%20of%20operation%20are%20similar.&hl=13:The%20PCCP%20does%20not%20include%20provisions%20for%20implementation%20of%20adaptive%20algorithms%20that%20will%20continuously%20learn%20in%20the%20field.&focus=2 , K261085 https://fda.innolitics.com/device/K261085?doc=summary&hl=10:Rapid%20VO%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20the%20CT%20Angiography%20%28CTA%29 , K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field.&focus=2 , K262301 https://fda.innolitics.com/device/K262301?doc=summary&hl=9:Rapid%20ICH%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20non-enhanced%20head%20CT plus K253265 2025 and K262765 2026 . Viewed longitudinally, CADt authorizations fall into three architectural eras reflecting both how imaging models evolved and which regulatory templates FDA’s radiological imaging review teams had established precedent to review. The CADt category opened on February 13, 2018 with Viz.ai’s De Novo authorization for ContaCT DEN170073 https://fda.innolitics.com/device/DEN170073?doc=decision summary&hl=1:ContaCT%20is%20a%20notification-only%2C%20parallel%20workflow%20tool%20for%20use%20by%20hospital%20networks%20and%20trained%20clinicians%20to%20identify%20and%20communicate%20images%20of%20specific%20patients%20to%20a%20specialist , authorized in 137 days for a single finding: suspected large vessel occlusion LVO on head CTA. Six months later, Aidoc cleared the category’s first 510 k , BriefCase K180647 https://fda.innolitics.com/device/K180647?doc=summary&hl=3:indicated%20for%20use%20in%20the%20analysis%20of%20non-enhanced%20head%20CT%20images.%20The%20device%20is%20intended%20to%20assist%20hospital%20networks%20and%20trained%20radiologists%20in%20workflow%20triage%20by%20flagging%20and%20communication%20of%20suspected%20positive%20findings%20of%20pathologies%20in%20head%20CT%20images%2C%20namely%20Intracranial%20Hemorrhage%20%28ICH%29 , in 142 days for intracranial hemorrhage ICH on non-contrast head CT. In March 2019, CureMetrix introduced QFM with cmTriage K183285 https://fda.innolitics.com/device/K183285?doc=summary&hl=3:cmTriage%20does%20not%20send%20a%20proactive%20alert%20directly%20to%20the%20radiologist.&hl=4:Radiological%20Computer-Assisted%20Prioritization%20Software , establishing passive worklist prioritization without proactive alerts. Through 2021, most CADt clearances validated one pathology at a time. Aidoc built the largest CADt portfolio in the market 30 primary-code clearances in total, including 26 through 2025 predominantly through sequential pathology-by-pathology clearances: ICH K180647 , PE added alongside ICH labeling K190072 , C-spine fracture K190896 , 56 days , LVO K192383 , intra-abdominal free gas K193298 , 205 days , incidental PE K201020 , 131 days , and rib fractures K202992 . When Aidoc filed a multi-predicate Special 510 k in December 2020 K203508 https://fda.innolitics.com/device/K203508?doc=summary&hl=12:Performance%20data%20was%20not%20needed%20to%20support%20this%20modification%20to%20BriefCase. , received November 30, 2020, cleared December 29, 2020 in 29 days , it broadened the intended user population across six previously cleared indications using separate Indications for Use forms and no new clinical performance testing. Early single-code multi-finding exceptions did exist during Era 1: Qure.ai’s qER K200921 https://fda.innolitics.com/device/K200921?doc=summary&hl=5:qER%20is%20an%20artificial%20intelligence%20software%20device%20that%20screens%20head%20CT%20scans%20for%20signs%20of%20ICH%2C%20mass%20effect%2C%20midline%20shift%20and%20cranial%20fractures , June 2020 cleared four head CT pathologies under QAS ICH, cranial fracture, mass effect, and midline shift with user-selectable toggles, Avicenna.ai cleared two-finding bundles K200855 CINA for ICH and LVO in June 2020; K210237 CINA CHEST for PE and aortic dissection in May 2021 , and Lunit cleared a two-finding QFM chest X-ray device K211733 , November 2021 . Still, narrow task-specific convolutional networks kept one-indication-at-a-time filings the default. Between 2022 and 2024, two-finding single-code pairs became common across chest X-ray pneumothorax plus pleural effusion in Qure.ai K230899 , Milvue K232410 , and VUNO K241439 and stroke CT ICH plus LVO in Avicenna K221716 and Brainomix K232496 , while several single-pathology neuro clearances enumerated multiple vascular segments K220709 or hemorrhage subtypes K221456 , K241719 under one aggregate triage target. Annalise-AI later Harrison-AI Medical expanded the regulatory template further across 2022 and 2023. After clearing a two-finding QFM chest X-ray device for pneumothorax and tension pneumothorax K213941 https://fda.innolitics.com/device/K213941?doc=summary&hl=3:The%20device%20is%20designed%20to%20aid%20the%20clinical%20assessment%20of%20adult%20chest%20x-ray%20cases%20with%20features%20suggestive%20of%20pneumothorax%20and%20tension%20pneumothorax%20in%20the%20medical%20care%20environment.&hl=8:an%20AUC%20of%200.979%20%2895%25%20CI%3A%200.970-0.986%29%20and%200.988%20%2895%25%20CI%3A%200.981-0.993%29%20for%20pneumothorax%20and%20tension%20pneumothorax , February 2022, 70 days , Annalise established two precedents in consecutive weeks in spring 2023: Pioneering those two templates took calendar time: K222179 required 249 days and K223240 required 165 days . Once both summaries entered the public record, later sponsors had clear substantial-equivalence blueprints for dual-code mapping and unified multi-finding panels. In April 2025, Annalise became the first sponsor in the 132-record cohort to cross four distinct validated findings in one CADt 510 k , consolidating five chest X-ray findings under dual codes in K250831 https://fda.innolitics.com/device/K250831?doc=summary&hl=6:Note%20that%20product%20code%20QAS%20only%20applies%20to%20clinical%20condition%20of%20interest%3A%20pneumoperitoneum.%20All%20other%20conditions%20are%20submitted%20under%20product%20code%20QFM. a 35-day Special 510 k supported by a 3,252-case standalone evaluation across four previously acquired cohorts, mapping pneumoperitoneum to QAS and pleural effusion, pneumothorax, tension pneumothorax, and vertebral compression fracture to QFM . From late 2025 through November 2026, two distinct multi-indication scaling playbooks emerged. Between November 24, 2025 and November 12, 2026 353 calendar days between first and third clearance , a2z Radiology AI cleared three dual-code QAS + QFM , PCCP-authorized devices for abdominopelvic CT: a2z-Unified-Triage , cleared Nov 24, 2025, 117 days, PCCP : K223240 as its sole predicate, arguing that a2z-Abdo-Triage , cleared Sep 16, 2026, 181 days, PCCP : a2z-Omni-Triage , cleared Nov 12, 2026, 100 days, PCCP confirmed : Clinical overlap note: The sponsor’s reported 26 represents cumulative cleared finding claims across three separate devices rather than 26 mutually exclusive abdominal pathologies. In our clinical comparison of K252366 against the reported K262765 list, three of Omni’s nine claims refine severity, size, or etiology of finding families already cleared in K252366 moderate/severe hydronephrosis vs. hydronephrosis; aortic aneurysm = 5 cm vs. unruptured AAA; bowel perforation vs. free air and perforated diverticulitis , and three pleural effusion, pneumothorax, complex pleural collection are lower-thoracic pleuropulmonary findings visible at the lung bases on abdominopelvic CT. CARE Multi-Triage After 26 primary-code BriefCase clearances through 2025 and alongside two additional single-code clearances in early 2026 , Aidoc expanded into dual-code QAS + QFM multi-triage bundles built on a foundation model, covering patients aged 18 and older compared with 22 and older for a2z and Annalise : BriefCase-Triage: CARE Multi-triage CT Body , cleared Jan 7, 2026, 112 days : BriefCase-Triage: CARE Multi-Triage CT , cleared Feb 26, 2026, 101 days : | Submission | Decision Date | Sponsor & Device | Codes | Findings | Review Days | Validation Cohort Design | PCCP | |---|---|---|---|---|---|---|---| | DEN170073 https://fda.innolitics.com/device/DEN170073?doc=decision summary&hl=1:ContaCT%20is%20a%20notification-only%2C%20parallel%20workflow%20tool%20for%20use%20by%20hospital%20networks%20and%20trained%20clinicians%20to%20identify%20and%20communicate%20images%20of%20specific%20patients%20to%20a%20specialist | 2018-02-13 | Viz.ai ContaCT | QAS | 1 | 137 | 300 CTA cases category-creating De Novo | No | | K180647 https://fda.innolitics.com/device/K180647?doc=summary&hl=3:indicated%20for%20use%20in%20the%20analysis%20of%20non-enhanced%20head%20CT%20images.%20The%20device%20is%20intended%20to%20assist%20hospital%20networks%20and%20trained%20radiologists%20in%20workflow%20triage%20by%20flagging%20and%20communication%20of%20suspected%20positive%20findings%20of%20pathologies%20in%20head%20CT%20images%2C%20namely%20Intracranial%20Hemorrhage%20%28ICH%29 | 2018-08-01 | Aidoc BriefCase ICH | QAS | 1 | 142 | First CADt 510 k ; head CT standalone cohort | No | | K183285 https://fda.innolitics.com/device/K183285?doc=summary&hl=3:cmTriage%20does%20not%20send%20a%20proactive%20alert%20directly%20to%20the%20radiologist.&hl=4:Radiological%20Computer-Assisted%20Prioritization%20Software | 2019-03-08 | CureMetrix cmTriage | QFM | 1 | 102 | First passive worklist QFM clearance | No | | K203508 https://fda.innolitics.com/device/K203508?doc=summary&hl=12:Performance%20data%20was%20not%20needed%20to%20support%20this%20modification%20to%20BriefCase. | 2020-12-29 | Aidoc BriefCase | QAS | 6 prior | 29 | Special 510 k user-label update; no new clinical testing | No | | K222179 https://fda.innolitics.com/device/K222179?doc=summary&hl=4:Note%20that%20the%20product%20code%20QFM%20applies%20to%20pleural%20effusion.&hl=4:Note%20that%20the%20product%20code%20QAS%20applies%20to%20pneumoperitoneum.&hl=8:total%20dataset%20of%201269%20CXR%20cases%20%28positive%20n%3D582%20and%20negative%20n%3D687%29%2C%20collected%20consecutively%20from%20four%20US%20hospital%20network%20sites. | 2023-03-28 | Annalise CXR Trauma | QFM + QAS | 2 | 249 | 1,269 CXRs across 4 U.S. hospital network sites | No | | K223240 https://fda.innolitics.com/device/K223240?doc=summary&hl=3:These%20findings%20are%20intended%20to%20be%20used%20together%20as%20one%20device. | 2023-04-03 | Annalise CTB Trauma | QAS | 4 | 165 | 1,485 thin + 1,878 thick CTBs, 5 U.S. network sites | No | | K250831 https://fda.innolitics.com/device/K250831?doc=summary&hl=6:Note%20that%20product%20code%20QAS%20only%20applies%20to%20clinical%20condition%20of%20interest%3A%20pneumoperitoneum.%20All%20other%20conditions%20are%20submitted%20under%20product%20code%20QFM. | 2025-04-23 | Annalise Enterprise CXR | QFM + QAS | 5 | 35 | Special 510 k ; 3,252 CXRs across 4 cohorts | No | | K252366 https://fda.innolitics.com/device/K252366?doc=summary&hl=7:While%20the%20anatomical%20regions%20of%20interest%20differ%20%28abdomen%2Fpelvis%20for%20the%20subject%20device%20vs.%20head%20for%20the%20predicate%29%2C%20the%20core%20technology%2C%20intended%20use%20for%20triage%2C%20and%20principles%20of%20operation%20are%20similar.&hl=13:The%20PCCP%20does%20not%20include%20provisions%20for%20implementation%20of%20adaptive%20algorithms%20that%20will%20continuously%20learn%20in%20the%20field. | 2025-11-24 | a2z-Unified-Triage | QAS + QFM | 7 | 117 | 675 analyzed 678 reviewed, 643 patients , 12 sites, 2+1 | Yes | | K252970 https://fda.innolitics.com/device/K252970?doc=summary&hl=11:Multi-triage%20module%2C%20locked%20artificial%20intelligence%20algorithm%20fine%20tuned%20from%20a%20foundation%20model.&hl=12:Each%20of%2011-clinical%20indications%20had%20a%20sample%20size%20of%20N%20%3D%20280%2C%20with%201769%20unique%20scans%20included%20across%20device%20indications. | 2026-01-07 | Aidoc CARE CT Body | QAS + QFM | 11 | 112 | N = 280 per finding, 1,769 unique scans, 6 sites | No | | K253578 https://fda.innolitics.com/device/K253578?doc=summary&hl=10:Each%20of%204-clinical%20indications%20had%20a%20sample%20size%20of%20N%20%3D%20280%20each%2C%20with%20772%20unique%20scans%20included%20across%20device%20indications. | 2026-02-26 | Aidoc CARE Multi-Triage CT | QAS + QFM | 4 | 101 | N = 280 per finding, 772 unique scans, 6 sites | No | | K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field. | 2026-09-16 | a2z-Abdo-Triage | QAS + QFM | 10 | 181 | 965 analyzed 989 reviewed, 914 pseudonyms , 3 sites, 2+1 | Yes | | K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 | 2026-11-12 | a2z-Omni-Triage | QAS + QFM | 9 | 100 | Summary PDF not yet posted in FDA’s 510 k database | Yes | Nine findings per ITN Nov 2026 https://highlight.innolitics.com/v/jLF0v46yMMDH h=2 . Primary QAS, subsequent QFM https://highlight.innolitics.com/v/pp1p85c14Mlw h=2 , and PCCP Authorized: Yes https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 confirmed via Innolitics FDA Device Explorer https://fda.innolitics.com/device/K262765 and FDA’s 510 k record; PCCP scope remains unverified until the summary posts. Comparing the 2025 and 2026 summaries highlights four concrete architectural choices for teams planning multi-indication CADt submissions. The two scaling sponsors took distinct statistical approaches to standalone validation: N = 280 cohort for each indication across 6 U.S. sites, drawn from a larger unique scan pool 1,769 and 772 unique scans, respectively . Both summaries report sensitivity and specificity across default and four additional operating points alongside AUC, with different stated primary AUC goals: AUC 0.95 , whereas 0.95 for AUC. Aligning on cohort structure, enrichment ratios, primary acceptance bounds, and reader adjudication rules in a voluntary Pre-Submission Q-Sub reduces the risk of mid-review study redesign. Both scaling sponsors in 2025 and 2026 grouped findings into modular waves 7, 10, and 9 claims for a2z; 11 and 4 for Aidoc CARE , and across all 132 primary-code CADt clearances no sponsor has cleared a 15-plus-finding mega-bundle. While public clearance records cannot reveal private sponsor motives or prove fewer holds, three planning considerations explain why modular waves are practical: 117d for 7 findings, 181d for 10, 100d for 9 are consistent with manageable complexity at the 4-to-11-finding scale, though public records do not disclose whether | Regulatory Strategy | Findings / Filing | Observed Review Examples | Planning Strengths | Key Planning Trade-Off | |---|---|---|---|---| | Single-Finding v1 Era 1 baseline | 1 | 56 to 205 days varies widely | Narrowest validation and review scope | Repeats user fees and QMS overhead per finding; risks hospital “point-solution fatigue” | | Modular Sprint Observed 2025-26 Pattern: a2z 7/10/9 , Aidoc CARE 11/4 | 4 to 11 | 100 to 181 days Traditional ; 35 days K250831 Special | Shared-cohort or scan-pool reuse; dual QAS + QFM coverage; staggered filings separate hold exposure | Requires disciplined multi-label ground-truthing and upfront QAS/QFM workflow alignment | | Monolithic Mega-Bundle 0 clearances across all 132 records | 15 to 25+ | No precedent in 21 CFR 892.2080 | Single user fee on paper | One weak rare finding can pause the entire file; heavy annotation spend before first revenue | Teams scoping a multi-finding triage device sometimes assume QAS vs. QFM maps cleanly to clinical acuity or alert routing alone. Published clearances show that alert routing alone cannot predict the product code: K213941 = 80% at cleared operating points. Confirming each finding’s product-code mapping and both AUC and operating-point acceptance goals in a voluntary Q-Sub prevents mid-review misalignment. Across the 132 primary-code authorizations, at least 6 carry an authorized PCCP: two in 2025 Aidoc’s Special 510 k K253265 https://highlight.innolitics.com/v/3r61d7uINGlI h=1 and a2z’s K252366 https://fda.innolitics.com/device/K252366?doc=summary&hl=7:While%20the%20anatomical%20regions%20of%20interest%20differ%20%28abdomen%2Fpelvis%20for%20the%20subject%20device%20vs.%20head%20for%20the%20predicate%29%2C%20the%20core%20technology%2C%20intended%20use%20for%20triage%2C%20and%20principles%20of%20operation%20are%20similar.&hl=13:The%20PCCP%20does%20not%20include%20provisions%20for%20implementation%20of%20adaptive%20algorithms%20that%20will%20continuously%20learn%20in%20the%20field.&focus=2 and four in 2026 iSchemaView’s K261085 https://fda.innolitics.com/device/K261085?doc=summary&hl=10:Rapid%20VO%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20the%20CT%20Angiography%20%28CTA%29 Rapid Vessel Occlusion and K262301 https://fda.innolitics.com/device/K262301?doc=summary&hl=9:Rapid%20ICH%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20non-enhanced%20head%20CT Rapid ICH v4.0, which enumerates four hemorrhage subtypes under an aggregate ICH endpoint, plus a2z’s K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field.&focus=2 and K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 . Founders often ask whether an authorized PCCP lets them add net-new pathologies to a cleared triage model without a new 510 k . None of the four published CADt PCCP summaries examined include new clinical findings. The PCCP scopes of K253265 https://highlight.innolitics.com/v/3r61d7uINGlI h=1 and K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 are not detailed in posted summaries. Both published a2z summaries K252366 https://fda.innolitics.com/device/K252366?doc=summary&hl=7:While%20the%20anatomical%20regions%20of%20interest%20differ%20%28abdomen%2Fpelvis%20for%20the%20subject%20device%20vs.%20head%20for%20the%20predicate%29%2C%20the%20core%20technology%2C%20intended%20use%20for%20triage%2C%20and%20principles%20of%20operation%20are%20similar.&hl=13:The%20PCCP%20does%20not%20include%20provisions%20for%20implementation%20of%20adaptive%20algorithms%20that%20will%20continuously%20learn%20in%20the%20field.&focus=2 and K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field.&focus=2 exclude continuous field adaptation and scope planned retraining, architecture refinements, and operating-point updates to the conditions already cleared in each submission, while iSchemaView’s two PCCPs K261085 https://fda.innolitics.com/device/K261085?doc=summary&hl=10:Rapid%20VO%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20the%20CT%20Angiography%20%28CTA%29 and K262301 https://fda.innolitics.com/device/K262301?doc=summary&hl=9:Rapid%20ICH%20has%20an%20authorized%20Predetermined%20Change%20Control%20Plan%20%28PCCP%29.%20All%20modification%20types%20are%20restricted%20to%20non-enhanced%20head%20CT cover device-input compatibility, scanner expansion, performance improvement, and processing-time optimization. Under FD&C Act § 515C b and FDA’s PCCP guidance https://fda.innolitics.com/guidance/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence?v=gfb23b96999e-80f0f423efe0c810&hl=26:Modifications%20included%20in%20a%20PCCP%20must%20maintain%20the%20device%20within%20the%20device%27s%20intended%20use , modifications must maintain the device’s intended use, and indication-for-use changes in a PCCP are treated narrowly; teams planning net-new clinical pathologies should plan on a new 510 k under 21 CFR 807.81 a 3 . There is no statutory cap under 21 CFR 892.2080 . Across all 132 primary-code CADt authorizations in our Innolitics FDA Device Explorer census, FDA has cleared up to 11 distinct clinical findings in a single 510 k K252970 https://fda.innolitics.com/device/K252970?doc=summary&hl=11:Multi-triage%20module%2C%20locked%20artificial%20intelligence%20algorithm%20fine%20tuned%20from%20a%20foundation%20model.&hl=12:Each%20of%2011-clinical%20indications%20had%20a%20sample%20size%20of%20N%20%3D%20280%2C%20with%201769%20unique%20scans%20included%20across%20device%20indications.&focus=2 , Aidoc CARE Multi-triage CT Body and 10 findings in a single shared-cohort abdominopelvic CT submission K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field. , a2z-Abdo-Triage . In 2025 and 2026, scaling sponsors cleared 4 to 11 findings per filing ; future teams choosing a bundle size must weigh cohort reuse against annotation cost and hold exposure. Both are Class II special-controls codes under 21 CFR 892.2080 . FDA classifies QAS https://highlight.innolitics.com/v/maV4mtmNuRN2 h=1 as radiological computer-aided triage and notification software and QFM https://highlight.innolitics.com/v/AmyiUigQF7PE h=1 as radiological computer-assisted prioritization software for lesions first cleared with K183285 https://fda.innolitics.com/device/K183285?doc=summary&hl=3:cmTriage%20does%20not%20send%20a%20proactive%20alert%20directly%20to%20the%20radiologist.&hl=4:Radiological%20Computer-Assisted%20Prioritization%20Software for passive PACS worklist flags without proactive alerts . Alert routing alone cannot predict the assigned code: QAS-only K223240 https://fda.innolitics.com/device/K223240?doc=summary&hl=3:These%20findings%20are%20intended%20to%20be%20used%20together%20as%20one%20device. and dual-code K250831 https://fda.innolitics.com/device/K250831?doc=summary&hl=6:Note%20that%20product%20code%20QAS%20only%20applies%20to%20clinical%20condition%20of%20interest%3A%20pneumoperitoneum.%20All%20other%20conditions%20are%20submitted%20under%20product%20code%20QFM. both route study-level outputs to PACS/RIS worklists, and K250831 assigns pneumoperitoneum to QAS while placing four other CXR findings including tension pneumothorax under QFM. Multi-finding submissions increasingly combine QAS and QFM, so sponsors should consider workflow labeling, clinical urgency, and predicate lineage early and confirm each finding’s code and validation criteria including both AUC and sensitivity/specificity operating-point goals with FDA. None of the four published CADt PCCP summaries examined cover adding net-new clinical conditions; they govern bounded locked-model updates retraining, architecture refinements, operating-point adjustments, or input/scanner expansion within already-cleared conditions. Because FDA guidance https://fda.innolitics.com/guidance/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence?v=gfb23b96999e-80f0f423efe0c810&hl=26:Modifications%20included%20in%20a%20PCCP%20must%20maintain%20the%20device%20within%20the%20device%27s%20intended%20use treats indication-for-use changes in a PCCP narrowly and 21 CFR 807.81 a 3 requires a new 510 k for major intended-use changes or modifications that could significantly affect safety or effectiveness, plan on a new 510 k when adding new pathologies. Across all 132 primary-code CADt authorizations since 2018, the median total calendar review time receipt to decision is 113.0 days . Observed review times for recent multi-finding dual-code bundles span 35 days for Annalise’s Special 510 k K250831 https://fda.innolitics.com/device/K250831?doc=summary&hl=6:Note%20that%20product%20code%20QAS%20only%20applies%20to%20clinical%20condition%20of%20interest%3A%20pneumoperitoneum.%20All%20other%20conditions%20are%20submitted%20under%20product%20code%20QFM. and 100 to 181 days across five Traditional 510 k s K262765 https://highlight.innolitics.com/v/pp1p85c14Mlw h=1 : 100d, K253578 https://fda.innolitics.com/device/K253578?doc=summary&hl=10:Each%20of%204-clinical%20indications%20had%20a%20sample%20size%20of%20N%20%3D%20280%20each%2C%20with%20772%20unique%20scans%20included%20across%20device%20indications. : 101d, K252970 https://fda.innolitics.com/device/K252970?doc=summary&hl=11:Multi-triage%20module%2C%20locked%20artificial%20intelligence%20algorithm%20fine%20tuned%20from%20a%20foundation%20model.&hl=12:Each%20of%2011-clinical%20indications%20had%20a%20sample%20size%20of%20N%20%3D%20280%2C%20with%201769%20unique%20scans%20included%20across%20device%20indications. : 112d, K252366 https://fda.innolitics.com/device/K252366?doc=summary&hl=7:While%20the%20anatomical%20regions%20of%20interest%20differ%20%28abdomen%2Fpelvis%20for%20the%20subject%20device%20vs.%20head%20for%20the%20predicate%29%2C%20the%20core%20technology%2C%20intended%20use%20for%20triage%2C%20and%20principles%20of%20operation%20are%20similar.&hl=13:The%20PCCP%20does%20not%20include%20provisions%20for%20implementation%20of%20adaptive%20algorithms%20that%20will%20continuously%20learn%20in%20the%20field. : 117d, K260906 https://fda.innolitics.com/device/K260906?doc=summary&hl=13:The%20standalone%20performance%20assessment%20validated%2010%20abdominopelvic%20findings%20comprising%203%20QFM%20findings%20%28Hydroureter%2C%20Large%20volume%20ascites%2C%20Obstructing%20ureteral%20stone%29%20and%207%20QAS%20findings&hl=13:The%20PCCP%20does%20not%20authorize%20adaptive%20algorithms%20that%20continuously%20learn%20in%20the%20field. : 181d . Scoping a multi-finding triage submission is as much an architectural and biostatistical decision as a regulatory one. Choosing the right bundle size, predicate chain, cohort enrichment plan, and QAS/QFM workflow split early can save months of rework. At Innolitics, our medical imaging engineers, regulatory specialists, and clinical study leads help radiology AI teams scope multi-indication 510 k strategies, design voluntary Q-Submissions and multi-label validation protocols, author PCCPs, and build FDA-ready SaMD. Talk with our team https://innolitics.com/contact/ to map out your next CADt filing. Download the 1-page landscape brief with 34 clickable links to primary FDA summaries, guidance, and source reports: From Single-Finding Filings to Multi-Indication CADt 2018-2026 PDF https://innolitics.com/pdf/cadt-evolution-landscape-1pager.pdf .